1
Wavecom Instruments
Portable Appliance Testers
Instruction Manual
Australian made to comply with the requirements of the Standard AS/NZS 3760
Developed By Wavecom Instruments Pty Ltd
Ph: (08) 8243 3500
257 Grange Road
Findon SA 5023
Email: winpats@wavecom.com.au
Web: www.wavecom.com.au
Melbourne Office
772A Station Street, Box Hill, Victoria 3128
Phone: 1300 793 301
Fax: (+61) 03 9897 4766
Email: salesvic@wavecom.com.au
Vedere la pagina 0
1 2 3 4 5 6 ... 35 36

Sommario

Pagina 1 - Instruction Manual

1 Wavecom Instruments Portable Appliance Testers Instruction Manual Australian made to comply with the requirements of the Standard AS/NZS 37

Pagina 2 - Table of Contents

10 Earthed Appliance Test - F1 (For earthed appliances CLASS 1) Please Note: A visual inspection test must be carried out before any others (refer t

Pagina 3

11 5. Read and record results appropriately 6. Unplug DUT 7. Press the enter key to return to the main menu If the result was a pass - Tag the

Pagina 4 - Important Information

12 TnT-el Part No: WCM-TNT-EL TnT-el Test Function Flow Chart Menu 1 Note: Menu 1 is common to all models Menu 2 Main Menu 1 F1 Earth Test F2 D

Pagina 5 - Competent Person

13 Leakage Test The Leakage Test is an alternate method to perform insulation resistance tests. Leakage testing is a major function of the TnT-el &am

Pagina 6 - Technical Information

14 Caution: This test turns the Device Under Test (DUT) ON. The power / leakage test can handle up to 10 Amp (Resistive) Load MAX. Exceeding this

Pagina 7

15 Tests Common To All Models - Menu 1 Earthed Appliance Test (refer to TnT) Double Insulation Test (refer to TnT) Ext Lead Test (refer to TnT) Powe

Pagina 8 - Integrated Tests

16 Note: Before operating, ensure the equipment firmly secured and is free from any possibility of causing injury or damage. This function turns the

Pagina 9 - Visual Inspection

17 To select the second menu, press and release the Return/Enter button. This menu will display 3 functions:  Power Test  Leakage Test  RCD

Pagina 10 - Double Insulation Test - F2

18 Performing a Time Test: The displayed trip time is in milliseconds. This is the time taken for the RCD device to trip once the injected fault curr

Pagina 11 - Ext Lead Test - F3

19 RCD Timed Test (continued): Press the F1 key to select the 0 deg test and the F2 key to select the 180 deg test. Use the F1/F2 keys to scroll th

Pagina 12

2 Table of Contents Important Information  Accessories Supplied 4  Manufacturer Recommendations 4  Safety Warning 4  Disclaimer & L

Pagina 13 - Leakage Test

20 TnT-elx Part No: WCM-TNT-elx-15 (15A Plug) Part No: WCM-TNT-elx-20 (20A Plug) Before continuing reading this manual it is expected that you have

Pagina 14

21 TnT+X Part No: WCM-TNT+X-15 (15A Plug) Part No: WCM-TNT+X-20 (20A Plug) Before continuing reading this manual it is expected that you have the

Pagina 15 - Power Test - Menu 2

22 3PL Part No: WCM-TNT-3PL Before continuing reading this manual it is expected that you have read the previous sections. This includes reading all

Pagina 16 - RCD Test - Menu 2

23 Power Up Checking When the 3PL is turned on it will determine the type power source. This is either: - 1 Phase 3 Wire - 3 Phase 4 Wire (No Neutr

Pagina 17

24 These modes are displayed when the 3PL is first powered on. If a test is not available based on the type of power input then a message will be dis

Pagina 18

25 Tests Common to 3PL & TnT+ Models Menu 1 Earthed Appliance Test (refer to TnT) Double Insulation Test (refer to TnT) Ext Lead Test (refer to T

Pagina 19

26 Adding New Equipment The following must be done before testing:  Create a site or use and existing site.  Create a user name to under the user

Pagina 20 - TnT-elx

27 Edit User X ALP Entering a user is the same process as entering a site, please refer to the Adding a Site steps above, Note: The user

Pagina 21 - Tests Common To All Models

28 Main Menu A Earthed Test D/Insulated Test Ext Lead Test If the item passes a visual inspection and pass is selected, testing can begin. Sel

Pagina 22 - 3PL Diagram

29 Deleting Equipment If for some reason a record needs to be deleted it can be removed from the search barcode screen. This feature removes the wh

Pagina 23 - Menu System

3  Stop 21  TnT+X Diagram 21  TnT+X Test Function Flow Chart 21  Tests Common To All Models 21 3PL Instructions  Stop 22  3PL Diagr

Pagina 24

30 T&P Flow Chart Options Menu TnT Options Here you can adjust the way the TnT behaves under certain tests and other TnT functions. Some TnT o

Pagina 25 - TnT Memory Operation

31 You should now be in the TnT options screen. The screen displays 1 option at a time.  You can select this option by pressing the return button

Pagina 26 - Adding New Equipment

32  Press the key to disable.  Press the key to enable. Label Options - Only El Memory & Plus Memory Models Changing the label options ch

Pagina 27

33  Press the key to enable.  Press the key to disable. Barcode Options - Only El, Plus & 3PL Memory Models Barcode options refers to mod

Pagina 28 - Retesting Equipment

34 Optional Accessories WinPATS LITE Software Part No: WCM-WinPATS Lite WinPATS Base Software Part No: WCM-WinPATS Base WinPATS Base Software Upgrade

Pagina 29 - Test & Print (T&P)

35 Power Test (for TnT+ model only) ___ ________  Voltage 220 – 270 VAC + < 5% of reading  C

Pagina 30 - Options Menu

36 Relevant Authorities WorkCover NSW GPO Box 5364 Sydney NSW 2001-0423 Australia Ph: 02 4321 5000 Dept. of Fair Trading – NSW Safety and Standar

Pagina 31

4 Important Information Accessories Supplied (With the Range of TnT Portable Appliance Testers)  Portable Appliance Tester Instruction Manual  M

Pagina 32

5 TnT Portable Appliance Tester Competent Person To ensure that all electrical equipment or devices that are used to perform specific tasks, are ins

Pagina 33 - Other Features

6 Disclaimer Limited Warranty The Manufacturer warrants its products against defects in materials and workmanship for a period of 12 months from th

Pagina 34 - Specifications

7 Class II equipment is marked with the words "DOUBLE INSULATION" or the symbol For double insulated under Safety Symbols: Note 1 – Doubl

Pagina 35

8 To Change Test Voltage To change the test voltages select the "change ins" from the options menu. Please refer to the special functions s

Pagina 36 - Relevant Authorities

9 Visual Inspection (First Test) Visual Inspection THIS TEST IS TO OCCUR BEFORE ANY OTHER TEST IS CARRIED OUT USING ANY OF THE RANGE OF TnT APPLIANCE